NAVIGATION: Closed-loop control > Torque control > Current/torque controller settings

Current/torque controller settings

Current controller optimization

The torque controller is executed as a PI controller. The gain (P-component) and the integral-action time (I-component) of the individual controllers are programmable.
In order to optimize the current control loop, two rectangular reference steps are preset. The object of the optimization is a current controller with moderate dynamism and the following values:

Bild: Optimization of speed control


Vorgehensweise: Current controller optimization

The faster the actual value approaches the setpoint (reference), the more dynamic is the controller setting. During settling, the overshoot of the actual value should be no more than 5-10% of the reference (guide value).


Bild: Current controller optimization


Creating the transfer function

The oscilloscope automatically records the amount and phase response of the controller according to the controller settings. This produces an initial estimate of the control quality.

To determine the transfer function the noise amplitude (motor rated current) and the sampling time (default 0.125 ms) must be specified. Click the "Start Test Signal" button.

 

Bild: Noise amplitude, sampling time





Bild: Current controller transfer function


Green curve = Amount
Y-axis left = Absolute value of isd/isdref
Blue curve = Phase response
Y-axis right = Phase response Isd / isdref